$4.79 (+shipping and tax for $7.23 total) Shenzhen Rongbo Jiachuang Technology Co, before the tariff was imposed on the Chinese goods.
The device draws about 22mA by itself. There is 0.2-Ohm resistance between the input and the output. The current measurements seem to have about -10mA offset, which is a large error in the low current measurements; above 200mA, the measurement error is less than 5%. It may have calibrated at around 0.5A, the common USB port current.
Fluke 87 | USB meter |
0.007 | 0.00 |
0.031 | 0.02 |
0.039 | 0.03 |
0.063 | 0.05 |
0.099 | 0.09 |
0.142 | 0.13 |
0.205 | 0.20 |
0.467 | 0.47 |
0.903 | 0.92 |
1.438 | 1.47 |
2.146 | 2.17 |
The voltage measurements have 30-40mV offset. And there is some sort of data manipulation, the voltage measurements do not update consistently.
It is a good enough for casual USB power measurements.